EasyManua.ls Logo

GE Vivid i n - Brightness Adjustment Procedure; Testing Your Contrast and Brightness Settings; Blue Tint Adjustment Procedure

GE Vivid i n
740 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GE HEALTHCARE
DIRECTION FQ091013, REVISION 1 VIVID I N AND VIVID Q N SERVICE MANUAL
Chapter 6 - Service Adjustments 6-3
To assist in making Contrast adjustments, an adjustment bar is displayed below the screen:
Note: It is highly recommended to set the contrast to maximum. Press <ALT>+<Steer Right> to
increase the number of green bars to 7 (maximum).
6-2-2 Brightness Adjustment Procedure
To adjust the Brightness: hold down <ALT>:
- Repeatedly press to decrease the brightness
- Repeatedly pressing to increase the brightness.
To assist in making Brightness adjustments, an adjustment bar is displayed below the screen:
Note: The screen Brightness is calibrated when the dark blocks on the left (numbers 1 through 6)
- see Figure 6-1 - and the white bars on the right (numbers 1 through 12) are all just visible.
6-2-3 Testing Your Contrast and Brightness Settings
Observe the V watermark on the lower part of the monitor beside the soft keys on the screen. The
monitor is set to optimal settings when you just start seeing the
V.
- If you do not see the
V, the monitor is set too dark.
-If the
V is clearly visible, the monitor has too much contrast or is set too bright.
The Test Pattern in the Calibration Screen (refer to Figure 6-1 on page 6-2) may be used for Advanced
Settings. (Press Alt + Q to access the Test Pattern).
6-2-4 Blue Tint Adjustment Procedure
This will make the white shades shift a bit towards bluish.
Figure 6-2 LCD Display - Contrast Adjustment
Figure 6-3 LCD Display - Brightness Adjustment
Figure 6-4 LCD Display - V Watermark Adjustment