GE HEALTHCARE
DIRECTION FQ091013, REVISION 1 VIVID I N AND VIVID Q N SERVICE MANUAL
Chapter 8 - Replacement Procedures 8-81
5) Reconnect the AC power cable to the AC input assembly at the base of the cart (rear) and turn ON
power to the SafeLock Cart.
6) Proceed to perform the following functionality tests:
- SafeLock Cart - Grounding Continuity on page 10 - 23
- SafeLock Cart - Chassis Current Leakage Test on page 10 - 25