DIRECTION FQ091013, REVISION 1 VIVID I N AND VIVID Q N SERVICE MANUAL
Chapter 4 - General Procedures and Functional Checks 4-27
4-3-1-9 Front End Test
4-3-1-10 CW Test
Table 4-13 Front End Test - Recommended
Step Task Expected Result(s)
1
Boot up the system in Maintenance Mode, using the
Service Dongle.
2
Press Alt + Config to enter Diagnostics Mode
3
From the Diagnostic Test tree, select Front End Test.
4
Run Full System Test and press Start PASS
5
Connect the 3S-RS probe to the probe port.
6
From the Diagnostic Test tree, select Front End Test.
7
Select TR32 then select TX. Select TXProbeTX Test
and press Start
PASS
8
Select MUX Test then select 3S-RS Test and press
Start
PASS
9
Select TR32 Test then select RX Test. Select RSTest
with 3S-RS and press Start
PASS
10
When done, press the Esc button
Table 4-14 CW Test - Recommended
Step Task Expected Result(s)
1
Boot-up the System in Regular Scanning Mode
2
Press Alt + Config to enter Diagnostics Mode
3
From the Diagnostic Test tree, select Front End Test.
4
Run Full System Test and press Start PASS
5
Connect the 3S-RS probe to the probe port.
6
From the Diagnostic Test tree, select Front End Test.
7
Select TR32 then select TX. Select TXProbeTX Test
and press Start
PASS
8
Select MUX Test then select 3S-RS Test and press
Start
PASS
9
Select TR32 Test then select RX Test. Select RSTest
with 3S-RS and press Start
PASS
10
When done, press the Esc button