Loading...
GE HEALTHCARE
DIRECTION FQ091013, REVISION 1 VIVID I N AND VIVID Q N SERVICE MANUAL
Chapter 8 - Replacement Procedures 8-129
9) Type in the password provided for the specific software, and approve by clicking OK.
10) Restart the system and commence scanning.
11.)Proceed to perform the following functionality tests:
- Software Configuration Checks on page 4 - 37
- Calibration Tests on page 7 - 25
- Full System Test on page 7 - 10
Install Software
Service Tools
Exit