DIRECTION FQ091013, REVISION 1 VIVID I N AND VIVID Q N SERVICE MANUAL
3-122 Section 3-10 - Connectivity Setup
6.) Select the IP-address down-arrow to choose the DICOM Query/Retrieve server from the pull-down
menu. In some cases, the server to use is the same as used for DICOM Storage.
If the server to use is missing from the list, select <Modify> from the pull-down menu and edit the
setup for one of the predefined servers.
7.) Enter the correct AE Title and Port Number for the DICOM Query/Retrieve server in the respective
fields in the Query/Retrieve screen.
Change Search Criterias
It is possible to set up special Search Criterias for DICOM Query/Retrieve. In most cases you may leave
the Search Criterias as is, and skip this adjustment.
Follow the steps below to change the Search Criterias parameters:
1.) Select Search Criterias
Figure 3-130 DICOM Query/Retrieve properties
IP-ADDRESS
DOWN-ARROW
AE TITLE
PORT NO