GE HEALTHCARE
DIRECTION FQ091013, REVISION 1 VIVID I N AND VIVID Q N SERVICE MANUAL
3-174 Section 3-10 - Connectivity Setup
c.) Click the Enable eVue button to activate the eVue operation.
The following message is displayed:
Click OK to close the message.
Note: The Disable eVue button is now active.
d.) When required, click the Disable eVue button to de-activate the eVue operation.
The following message is displayed:
NOTICE
IMPORTANT To avoid system configuration inconsistencies, before using the Enable eVue or
Disable eVue functions, make sure that your PC is connected to the appropriate network and that all
network cables are properly connected. Otherwise, configuration of your PC may become incorrect and
the Configuration utility will be unable to execute the Enable eVue or Disable eVue operations.
In cases of a peer-to-peer connection, before using Enable eVue always make sure the two computers
(Vivid i n/ Vivid q N scanner and your PC) are properly configured and connected directly.
Before using Disable eVue, your PC should be connected back to the local network.