DIRECTION FQ091013, REVISION 1 VIVID I N AND VIVID Q N SERVICE MANUAL
5-20 Section 5-3 - Front End Unit
Table 5-7 P & M Board Input DC Voltages
Input Description Connector - PIN# Connected From
12 V DC voltage, distributed in P&M Board J5-16..21 RFI Board
3.3 V DC voltage, distributed in P&M Board J5-5..8 RFI Board
-5 VA DC voltage, distributed in P&M Board
"A" indicates analog circuit
J5-13..15 RFI Board
5 VA DC voltage
"A" indicates analog circuit
J5-9..12 RFI Board
HVPSP Transmit high voltage varies from
+2.5 to +50 V
J5-2 RFI Board
HVPSN Transmit high voltage varies from
-2.5 to -50 V
J5-1 RFI Board