DIRECTION FQ091013, REVISION 1 VIVID I N AND VIVID Q N SERVICE MANUAL
Chapter 7 - Diagnostics/Troubleshooting 7-27
8.) Use the scroll arrows at the bottom of the Reports window (see Figure 7-8) to scroll through the
channels and view the corresponding results.
9) When done, trackball to the Close button and press Set to close the View Report window.
10) After reviewing the test results, proceed as follows:
If no problems were identified, it is recommended to re-boot the system (this is obligatory when
prompted).
If the test passed but the original problem persists, contact your local On-Line Center for
assistance.
If errors were identified but the results were very close to the upper or lower limits, repeat the
DC Offset Calibration test again and review the newly-acquired results.
If errors were identified (results exceeding acceptable upper or lower limits), verify that all
system cables and boards are connected (not misplaced) and functioning properly, as
described in - Components and Function (Theory).
If the problem persists, this indicates that either the specific board you tested is faulty or that
another board that supports the operation of that board is faulty. Continue trouble-shooting
down to the appropriate board. (If necessary, perform a specific test).
If the problem still persists, contact the GE Medical Systems On-Line Center for assistance.
11.)Trackball to the Exit button and press Set to close the View Test Log dialog box.
12.)Press the <Esc> button on the keyboard to close the Diagnostic Test Window.