GE HEALTHCARE
DIRECTION FQ091013, REVISION 1 VIVID I N AND VIVID Q N SERVICE MANUAL
7-60 Section 7-4 - Performing Front End (FE) Diagnostics
If the problem persists, this indicates that either the specific board you tested is faulty or that
another board that supports the operation of that board is faulty. Continue trouble-shooting
down to the appropriate board. (If necessary, perform a specific test).
If the problem still persists, contact the GE HEALTHCARE On-Line Center for assistance.
13.)Trackball to the Exit button and press Set to close the View Test Log dialog box.
14.)Press the <Esc> button on the keyboard to close the Diagnostic Test Window.