GE HEALTHCARE
DIRECTION FQ091013, REVISION 1 VIVID I N AND VIVID Q N SERVICE MANUAL
8-96 Section 8-4 - SafeLock Cart Components Replacement
8-4-20-4 Rear Wheel Installation Procedure
1) Place the new rear wheel on the SafeLock Cart chassis in the same position as that of the
previously-removed wheel.
2) Return and fasten the four bolts (previously removed - see Figure 8-107) and make sure that wheel
is secured firmly to the chassis.
3) Stand the SafeLock Cart upright.
4) Return the Vivid i n/ Vivid q N and all peripherals to their appropriate positions on the SafeLock Cart.
5) Re-connect the AC adapter and all cables and peripherals to the Vivid i n/Vivid q N portable
ultrasound scanner.
6) Re-connect mains power to the AC adapter.
7) Proceed to perform the following functionality tests:
- SafeLock Cart - Grounding Continuity on page 10 - 23
- SafeLock Cart - Chassis Current Leakage Test on page 10 - 25