GE HEALTHCARE
DIRECTION FQ091013, REVISION 1 VIVID I N AND VIVID Q N SERVICE MANUAL
10-26 Section 10-7 - Electrical Safety Tests
10-7-5-3 Data Sheet for Chassis Source Current Leakage
The Chassis Source Current Leakage Test passes when all readings measure less than the values shown
in Table 10-14. Record all data on the Vivid i n/ Vivid q N System Inspection Certificate
(see page 10-34).
Table 10-17 Typical Data Sheet for Chassis Source Current Leakage Test
Unit
Power
Tester
Polarity
Switch
Tester
Neutral or
Ground
Switch
Test 1
Probe
Connector
Test 2
Caster
Wheel
Test 3
CRT
Test 4
(Optional)
Test 5
(Optional)
Enter Name of tested peripheral here:
ON NORM OPEN
ON NORM CLOSED
ON REV OPEN
ON REV CLOSED
OFF NORM OPEN
OFF NORM CLOSED
OFF REV OPEN
OFF REV CLOSED