Fltness Menu
command is set to array
CORR:SOUR
ARR, the array chosen is the
1601 point correction set.
n CORRection:SOURce[i]?
Queries the source of correction.
n CORRection[:STATe] ON]OFF]l]O
Sets the switch on the user flatness correction feature. This is the
same as pressing FLTNESS ON/OFF on the front panel.
n
CORRectionSTATe?
Queries the condition of the internal switch.
H
CORRection:FLATness:POINts? [MAXimum]MINimum]
The above command returns information on how many
frequency-correction pairs were entered using the CORR: FLAT
command.
Analyzer: NONE
See
Also
(ALC),
(FLTNESS ON/OFF), List Menu
“Optimizing Swept CW Generator Performance” in Chapter
I
“Programming Typical Measurements” in Chapter 1
FLTNESS ON/OFF)
Function Group
Menu Map
Description
Programming Codes
See Also
POWER
5
This
hardkey
applies flatness correction to the swept CW generator
RF output. If no array has been created, pressing this key applies 0
dB of flatness correction at all points. The yellow LED above the
hardkey
lights when user flatness correction is on.
SCPI: CORRection[:STATe] ON]OFF]l]O
Analyzer: NONE
@El),
Fltness Menu
“Optimizing Swept CW Generator Performance” in Chapter 1
F-10 Operating and Programming Reference