SCPI
COMMAND SUMMARY
l
FREQuency:MANual
<num>[freq
suffix]lMAXimumlMINimumlUPlDOWN
l
FREquency:MANual?
[MAXimumlMINimuml
Sets and queries the manual frequency. This controls the output
frequency in swept manual mode. The limits are START and STOP.
*RST
valueisthe same as
FREQ:CENTER.See
FREQ:CENTERfor
more
information.
l
FRECjuency:MODE
FIXedlCWlSWEeplLIST
l
FREquency:MODE?
Sets and queries the switch that selects either swept, CW or list
operation.
“RST
value is CW.
l
FREUuency:MULTiplier
Cnum>lMAXimumlMINimum
l
FREQuency:MULTiplier?
[MAXimumlMINimum]
Sets and queries the frequency multiplier. The numeric value is
rounded to the nearest integer. This function changes mapping of
frequency parameters on input to and output from the swept CW
generator. Changing this does not affect the output frequency of
the swept CW generator. Only the displayed parameters and query
responses are changed.
The equation implied by this is :
Entered/displayed frequency
=
(Hardware
Freq
* Multiplier ) + Offset
After
*RST,
the value is 1.
l
FREQuency:MULTiplier:STATe
ONlOFFlllO
l
FREQuency:MULTiplier:STATe?
Queries and turns the frequency multiplier off and on.
After
*RST,
the setting is OFF.
l
FREQuency:OFFSet
<num>lMAXimumlMINimum
l
FREQuency:OFFSet?
[MAXimumlMINimum]
Sets and queries the frequency offset. This function changes the
mapping of the frequency parameters on input to and output from
the swept
CW
generator. Changing this does not affect the output
frequency of the swept CW generator. Only the displayed parameters
and query responses are changed. The equation implied by this is :
Entered/displayed frequency
=
(Hardware
Freq
* Multiplier ) + Offset
After
“RST,
the value is 0.
l
FREQuency:OFFSet:STATe
ONlOFFlllO
l
FREUuency:OFFSet:STATe?
Operating and Programming Reference
S-3 1