SCPI
COMMAND SUMMARY
LIST:MODE
AUTO
LIST:TRIGger:SOURce
How the list is played back.
IMMediate
Each new frequency point is
stepped to automatically,
after waiting the specified
DWELl
time.
AUTO BUS
Wait for a <GET> or *TRG
over the HP-IB before
advancing to the next
frequency in the list.
AUTO
EXTernal Wait for a signal to be
received on the external
input before advancing to
the next frequency in the
list.
MAN&
Don’t care
Only the list point
specified by
LIST:MANual
is played back.
*RST state is AUTO.
l
LIST[:POWer]
:CORRection
C<num>CDBl
lMAXimumlMINimum~l*8Ol
l
LIST
[
:
POWer]
:
CORRect
ion?
Sets and queries the list of correction levels that correspond to each
of the frequencies entered using the
LIST:FREQ
command. The
attenuator is not allowed to change during the list execution. The
number of parameters can be from 1 to 801.
After
*RST,
the value is 0.
l
LIST
[
: POWer] :
CORRection
:
POINts?
[MAXimum
I
MINimum]
Returns the number of correction points that have been entered into
the list array. After *RST returns a 1.
l
LIST:TRIGger:SOURce
IMMediatelBUSlEXTernal
l
LIST:TRIGger:SOURce?
Sets and queries the list point-to-point trigger source when in the
automatic list mode. See LIST
:MODE
and INIT: CONT for more details.
*RST state is
IMMediate.
l
MARKerCnl
:AMPLitude[:STATel
ONlOFFlllO
l
MARKer
[nl
: AMPLitude
[:
STATe] ?
Sets and queries the amplitude marker on/off switch. While [n] may
be used, there is really only a single switch for all the markers.
*RST value is OFF.
l
MARKer
[n]
: AMPLitude :
VALue
<num>
[DBI
I
MAXimum
I
MINimum
l
MARKer
[n]
: AMPLitude :
VALue?
[MAXimum
I
MINimum]
Sets and queries the value of the amplitude marker. While [n] may
S-34 Operating and Programming Reference