HT2000H_en_n.doc Page 106 of 167
Confirmation
It injects the same sample in both front and rear injectors (see paragraph 11
“Appendix A – Glossary”.
• Name: shows the name of the sequence step.
Tap “Exit” to exit from the “STEP X” screen.
If any parameters have been changed, a confirmation screen appears stating “Modifying sequence step
X”. Tap “SAVE” to save the modifications or “CANCEL” to exit without saving.
The display return to the “RUN: SEQUENCES” screen.
5.3.3 Sequence tools
Access sequence: RUN>SETTINGS>SEQUENCES >TOOLS
In the “SEQUENCE TOOLS” screen (to see how to reach this screen, see paragraph 5.3
“HT2000H/HT2000HT: Sequence“), the following options are available:
• Add New: adds a new sequence step;
• Copy as New: creates a new sequence step by copying parameters from an existing one;
• Reorder: changes the order of the sequence steps;
• Remove: removes a sequence step.
If you select “ADD NEW” the following message appears: “A new step has been added at position X”.
Tap “CONTINUE” to confirm.
If you select “COPY AS NEW”, first select the source step to copy, then the following message appears
“A new step has been copied at position X”. Tap “CONTINUE” to confirm.
If you select “REORDER”, you have to select the new step position (target position) and tap “MOVE”. A
screen appears: “Step X has been moved to position Y”. “Y” will be the new step position.
If you select “REMOVE”, you have to select the step to remove, and then a screen appears: “Step X
has been removed”. Tap “CONTINUE” to confirm.
Select “Exit” to exit from this screen and return to the “RUN: SEQUENCES” screen (see paragraph
5.3.1 “Sequence Menu “).