HT2000H_en_n.doc Page 92 of 167
High
The fan is active (high speed)
Very
High
The fan is active (very high speed)
Press “OK” to move the fan at the set speed. This screen also shows:
Touch sensor: it is “ON” if the vial locator has detected the presence on an object, or
“OFF” if it is not touching.
Shaker sensor: it is “ON” when the shaker is in the position suitable for vial loading and
unloading, it is “OFF” in all the other positions (shaker sensor is not available for
HT2100H).
Ready: shows the “Ready” signal status (see paragraph 4.2.4.4 “Setup configuration: Input
Output”);
Start: shows the “Start” signal status (see paragraph 4.2.4.4 “Setup configuration: Input
Output”). By pressing “OK”, this signal is activated for about 1 second.
Sync-out: shows the “Sync-out” signal status. Pressing “OK” this signal is activated.
Sync-in: show the “Sync-in” signal status.
Flush: shows the status of the flushing valve. It is “ON” when the valve is open and the gas
can flush the syringe, it is “OFF” when the valve is closed. Press “OK” to change the valve
status.
Select Exit to exit from the “Setup Manual Operation” screen and return to the “Setup” screen
(see paragraph 4 “Set Up operations“).
4.2.4 Setup: Configuration
Access sequence: RUN>SETTINGS>SETUP>CONFIGURATION
This is used to define some general settings of the autosampler. The information is shown on
multiple pages.
Use this screen to access the following sub menus:
Sound: changes the sound effects of the autosampler (see paragraph 4.2.4.1 ”Setup
configuration: Sound”);
Light: changes the syringe illumination settings (see paragraph 4.2.4.2 “Setup
configuration: Light“);
Run: changes the settings for the injection sequence (see paragraph 4.2.4.3 “Setup
configuration: Run “);
Input Output: changes the input/output parameters to/from the analyzer (see paragraph
4.2.4.4 “Setup configuration: Input Output”);