Select “Exit” to exit from this screen and return to the “RUN: METHOD X” screen (see paragraph
5.2.2 “Method X: Setting method parameters”).
5.2.2.5 Injection parameters
Access sequence: RUN>SETTINGS >METHODS>METHOD X>INJECTION
The following parameters can be viewed/edited:
• Injection speed: the syringe dispensation speed used during injection;
• Pre Injection Dwell: the waiting time of the syringe needle inside the injection port before
making the injection;
• Post Injection Dwell: is the waiting time of the syringe needle inside the injector after the
injection;
• Flush Time: syringe flush time after the injection.
Select “Exit” to exit from this screen and return to the “RUN: METHOD X” screen (see paragraph 5.2.2
“Method X: Setting method parameters”).
5.2.3 Method tools
Access sequence: RUN>SETTINGS>METHODS>TOOLS
In the “METHOD TOOLS” screen (see paragraph 5.2 “HT2000H/HT2000HT: Methods“), the following
options are available:
• Restore Default: used to load the default settings for a method;
• Copy Methods: used to copy the parameters from one method to another.
If you select “RESTORE DEFAULT”, you have to select the method to reset. A confirmation screen
appears “Loading default parameter for method X”. Tap “LOAD” to restore the default parameters and
“CANCEL” not to make the restore operation. If you select “COPY METHODS”, you have to select the
source method and the destination method. Tap “CONTINUE” to confirm and “ABORT” not to make the
method copy.
Select “Exit” to exit from this screen and return to the “RUN: METHODS” screen (see paragraph
5.2.1 “Method Menu”).
5.2.4 Example
Below there is an example method, and you can use this to create your own method.
Please keep in mind that the example method should be adapted depending on your analysis
technique, analyzer model, inlet type, etc. and therefore this may not be suitable for your application.