HT2000H_en_n.doc Page 89 of 167
Plunger, syringe and needle are heated up to 150°C (standard version syringe) or 250°C (HT
version syringe installed on HT2000HT). We recommend taking care to prevent injury.
To remove the syringe warmer assembly follow the instructions given in paragraph 4.2.2 “Setup:
Syringe warmer assembly installation” excluding points “d)“-“f)“. Instead of executing these points
simply remove the syringe warmer assembly.
4.2.3 Setup: Manual operations
This function allows manual movement of the autosampler motors.
Access sequence: RUN>SETTINGS>SETUP>MANUAL OPERATION
This screen displays a list of the autosampler motors.