11 Appendix A – Glossary
Head Space Technique
Head Space technique is an indirect method to inject components with low boiling points from
liquid or solid samples. The method is based on the analysis of vapours released from a sample
that has been incubated at a constant temperature for a given period of time. Head Space
technique has some advantages compared to other techniques: quick preparation time compared
to analysis time, possibility to analyse samples that cannot be analysed in liquid or in solid state,
less cross contamination problems and higher sensitivity, as this procedure can be considered as a
pre-concentration step before injection.
Sequence
A sequence is an ordered series of sample vials to prepare and inject, including the method
needed to prepare each vial.
Confirmation Mode
Confirmation Mode Technique injects the same sample into two injectors to confirm the analysis.
The GC is started on the first injection.
Touch sensor
It is used by the autosampler to detect when the vial locator has touched a surface (or obstacle).
System integrity test
The System integrity kit is sold as an option (see paragraph 1.7.2 “Options “). The System integrity
test is used to verify the syringe integrity. To enable this function see paragraph 4.2.4.5 “Setup
configuration: Maintenance “.
If this test is enabled, at the beginning of each batch the system performs a check to verify the
syringe integrity through a heuristic procedure. This test is carried out using the system integrity
tool (see Figure 1 and Figure 2 for HT2000H/HT2000HT and Figure 9 for HT2100H).The septum
of this tool should be replaced periodically (see paragraph 3.6.3 “System integrity tool installation”).
This check can verify if it is better to replace the syringe (barrel and plunger). Please note that this
check does not replace the necessary periodic validations you have to program for the instrument
qualification. If the test fails, a warning message appears (see paragraph 8.14 “System integrity
test fail“).