HT2000H_en_n.doc Page 117 of 167
Figure 102: HT2000H/HT2000HT rack placing
7.6.2 Eco Savings
Warning
Only for HT2000HT. Standard version syringe installed in the Standard syringe warmer
assembly cannot work at temperature higher than 150°C. Do not set higher
temperatures.
This function is used to set the stand by temperatures. Access sequence: RUN>SETTINGS>ECO
SAVINGS In the “Eco Savings” screen you can set:
• St by syringe: sets the syringe stand by temperature. This value can be between 40°C and
150°C (standard version syringe)/ 250°C (HT version syringe installed on HT2000HT) or
off (inactive conditioning);
• St by oven: sets the oven stand by temperature. This value can be between 40°C
9
and
170°C (for HT2000H)/ 150°C (for HT2100H)/ 300°C (for HT2000HT) or off (inactive
conditioning);
9
Or the minimum oven temperature that can be set in your special autosampler version