HT2000H_en_n.doc Page 117 of 167
Figure 102: HT2000H/HT2000HT rack placing
7.6.2 Eco Savings
Warning
Only for HT2000HT. Standard version syringe installed in the Standard syringe warmer
assembly cannot work at temperature higher than 15C. Do not set higher
temperatures.
This function is used to set the stand by temperatures. Access sequence: RUN>SETTINGS>ECO
SAVINGS In the “Eco Savings” screen you can set:
St by syringe: sets the syringe stand by temperature. This value can be between 40°C and
15C (standard version syringe)/ 25C (HT version syringe installed on HT2000HT) or
off (inactive conditioning);
St by oven: sets the oven stand by temperature. This value can be between 40°C
9
and
17C (for HT2000H)/ 150°C (for HT2100H)/ 300°C (for HT2000HT) or off (inactive
conditioning);
9
Or the minimum oven temperature that can be set in your special autosampler version
G
row