HT2000H_en_n.doc Page 99 of 167
5 Programming
5.1 HT2100H: Methods and Sequence
HT2100H methods and sequences can only be programmed by PC, using the “HTA Autosampler
Manager”. After programming, the HT2100H may be disconnected from the PC and operated by
the integrated keypad as described in chapter 6 “HT2100H operations”.
5.2 HT2000H/HT2000HT: Methods
5.2.1 Method Menu
This function is used to edit or view the injection method parameters. Access sequence:
RUN>SETTINGS>METHODS
The “RUN: METHODS” screen contains a list of the 10 methods (from 0 to 9) stored in the unit
memory.
The methods can be identified by the number or by a name.
The name can only be created or edited using a PC (using the “HTA Autosampler Manager”
software). Once the names have been created, they will be visible on the touch screen.
Select a method number/name to enter in the “METHOD X” screen (X is a number from 0 to 9 that
can be followed by the method name) and to view or edit its parameters. See paragraph
5.2.2 “Method X: Setting method parameters”.
Select “Tools” to access the “METHOD TOOLS” screen (see paragraph 5.2.3 “Method tools “) Tap
“Exit” to exit from the “RUN: METHODS” screen.
5.2.2 Method X: Setting method parameters
Each method contains a set of analytical parameters. Use this screen to view or edit these
parameters.
Access sequence: RUN>SETTINGS>METHODS>METHOD X
In the “METHOD X” screen the following sub menus are available:
• General: see paragraph 5.2.2.1;
• Advanced: see paragraph 5.2.2.2;
• Preparation: see paragraph 5.2.2.3
• Sample: see paragraph 5.2.2.4