Tap “CONTINUE”. The “Setup: Service” screen appears (see paragraph 4.2.5 Setup: Service”).
Contact your Supplier or the Technical Assistance Centre to solve the problem.
8.2.2.2 Method reading error
Error message:
“Error during methods reading!"
Causes:
Autosampler cannot read the parameters of the saved methods.
Solution:
Tap “CONTINUE”. Try to switch on-off the autosampler. If the error message appears again,
contact your Supplier or the Technical Assistance Centre to solve the problem.
8.2.2.3 Sequence reading error
Error message:
“Error during sequences reading!"
Causes:
Autosampler cannot read the param.eters of the saved sequences.
Solution:
Tap “CONTINUE”. Try to switch on-off the autosampler. If the error message appears again,
contact your Supplier or the Technical Assistance Centre to solve the problem.
8.3 Method errors
8.3.1 Fill Volume/Pull Up Strokes settings incompatibility
Error message:
“Please add a volume for Fill Volume or set to 0 Pull-up strokes
Causes:
You have set a “Fill volume” to zero, but you have set “Pull Up Strokesto a value different to zero.
These settings are not compatible.
Solution:
Tap “CONTINUE” to edit the method (set “Fill volume” to a value different from zero or set the “Pull
Up Strokes” to zero).
8.3.1 Fill volume/Syringe Pre-Fill settings incompatibility
Error message:
“Please add a volume for Fill Volume or set to NO Syringe Pre-Fill”
Causes:
You have set a “Fill volume” to zero, but you have set the Syringe Pre-Fill to “YES”. These settings
are not compatible.