In the Run time mode, press the MOVE key to move the turret to allow free access to the
rack (see paragraph 6.1 “Keys“),
Clean the vial rack using a non-abrasive detergent solution.
Dry with a dry cloth.
Press the MOVE key to reposition the turret.
9.1.2 External cleaning of cabinet
Switch OFF the autosampler.
Remove the power cable from the rear.
Carefully clean the control panel (touch screen or keypad) and fragile parts using only a
damp cloth. They must not be cleaned with solvents or detergents.
Clean the cabinet using a non-abrasive detergent solution.
Dry with a dry cloth.
Do NOT use water, solvents, detergents or spray on electrical parts.
9.1.3 Incubation oven cleaning
Warning
Oven surface and spacers (when inserted) are heated at high temperature and can be very
hot. We recommend taking care in order to prevent burns.
Be sure that the autosampler is in stand-by status and that no other operations are in
progress.
Set the oven stand-by temperature to off (see paragraph 7.6.2 “Eco Savings”);
Only for HT2000H/HT2000HT: enter “manual operations” and open the oven cover (see
paragraph 4.2.3 “Setup: Manual operations”);
Turn OFF the autosampler and remove the power cable;
Wait until the incubator oven cools down;
Remove the spacers (if present) using the suitable tweezers ;
Clean the incubator and the spacers;
Reinstall the spacers (if needed);
Let the oven dry;
Connect the autosampler to the power supply and switch ON the unit;