HT2000H_en_n.doc Page 161 of 167
Syringe Temperature
Syringe Type
Off Standard version syringe
40-100°
Standard version syringe
100-150°C
Standard version syringe/
HT version syringe
151-250°C
Only HT version syringe
Warning
Only for HT2000HT. Standard version syringe installed in the Standard syringe warmer
assembly can not work at temperature higher than 150°C. Do not set higher temperatures. The
system can be seriously damaged.
12.2 Sample vials
HTA Part Number
Description
Temperature Range (°C)
1.29.949 6ml transparent vials with round end
1.29.941 10ml transparent vials with round end
1.29.942 20ml transparent vials with round end
1.29.940 Septa and caps for HS vials in SIL/PTFE
from -40 to +200°C.
12.2.1 Vial specifications
• Vials must be approved for head space techniques;
• Rounded end;
• Clear or amber glass (amber glass is suitable for light sensitive samples).