HT2000H_en_n.doc Page 98 of 167
tool (see Figure 1 and Figure 2 for HT2000H and HT2000HT and Figure 9 for HT2100H).The
septum of this tool should be replaced periodically (see paragraph 3.6.3 “System integrity tool
installation“).
This check can verify if it is better to replace the syringe (barrel and plunger). Please note that this
check does not replace the necessary periodic validations you have to program for the instrument
qualification. If the test fails, a warning message appears (see paragraph 8.14 “System integrity
test fail“).
Select “Exit” to exit from this screen and return to the “Setup Configuration”” screen (see
paragraph 4.2.4 “Setup: Configuration“).
4.2.4.6 Setup configuration: Advanced
Access sequence: RUN>SETTINGS>SETUP>CONFIGURATION>NEXT>ADVANCED
This menu is password protected and is only accessible to service representatives.
4.2.4.7 Setup configuration: Activation
Access sequence: RUN>SETTINGS>SETUP>CONFIGURATION>NEXT>Activation
This screen contains a list of 9 optional features (from 1 to 9). Select a feature to enable it (to know
if there feature available for activation, contact your sales representative).
A screen showing the following message: “You are activating the feature X” (“X” is the feature
number) appears.
• Tap CANCEL to return to the previous screen.
• Tap TEMPORARY to temporarily enable the selected feature (see below how to proceed).
• Tap FULL, to definitively enable the selected feature. A password is required for the
activation. Please contact your sales representative to get the password.
If you have tapped “TEMPORARY” a screen showing the following message: “Random number:
X” appears. If this is your first activation attempt, send this random number to your sales
representative to obtain the password for the temporary activation. Then tap “CONTINUE” to
access the screen for the password entry and temporarily enable the feature. If you haven't
received the password yet, from the screen for the password entry you can tap “CANCEL” and
retry the activation later. You have only to re-execute the described operations. The unit will re-
show you the same random number (because you have already tried the activation, but you
haven't complete it). Directly tap “ACTIVATE” to access the screen for the password entry and
temporarily enable the feature.
4.2.5 Setup: Service
Access sequence: RUN>SETTINGS>SETUP>SERVICE
This menu is password protected and is only accessible to service representatives.