EasyManua.ls Logo

HTA HT2000H - Syringe Location and Connection Panel

HTA HT2000H
167 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
HT2000H_en_n.doc Page 28 of 167
1
“X” marks the rubber feet, these are important for the installation procedure (see paragraph 3.3.1.1
“Unlock the autosampler from the mounting kit”).
2.1.2.5 HT2100H Control panel: keypad
In the front of HT2100H there is the keypad.
Figure 14: Keypad (HT2100H)
The HT2100H uses a keypad with 4 keys (Move, Start, Wash, Stop) and 4 status LEDs (Err,
Ready, Setup, Run). This keypad is used for sample execution and basic maintenance/setup tasks.
The method, sample list and set up editing is performed exclusively by a PC. For further
information on the HT2100H keypad, please refer to chapter 6 “HT2100H operations”.
2.1.3 Syringe location
The syringe is located in the turret. To access the syringe, push up the sliding lid (red arrow in the
figure below).
Figure 15: Sliding lid down Figure 16: Sliding lid up

Table of Contents

Related product manuals