HT2000H_en_n.doc Page 108 of 167
STOP should be pressed as soon as the error condition has been fixed.
o press once to retry the operation;
o press twice to abort current operations.
6.1.1 Keys Settings
HTA Autosampler Manager” can define the function of the “PURGE” and “START” keys.
In HTA Autosampler Manager”, select “Auxiliarytab and then “Key Settings”. If additional help is
needed, please refer toHTA Autosampler Manager” online help.
6.2 Status Indicators
On the keypad there are 4 LEDs: Err, Set, Run, Rdy.
6.2.1 Reference table
Legend:
OFF
ON
Blinking
Err
Set
Run
Rdy
Meaning
Autosampler is OFF. Press main switch on the back to power it ON.
Initialization in progress. Please wait for initialization to be completed.
Blocking error detected. Please connect “HTA Autosampler Manager”
and check “Virtual Screen” tab in order to get more information about
error type.
Autosampler is ready to accept new commands.
Autosampler is busy while executing a sample sequence or a single
wash.
Autosampler is in PAUSE status.
Press START to resume.
Press STOP to abort.
Autosampler is aborting operations in progress.