HT2000H_en_n.doc Page 109 of 167
Err
Set
Run
Rdy
Meaning
Autosampler is in SETUP mode.
SETUP mode may be entered in an automatic way (error detected) or in
a manual way (by PC software).
Please connect “HTA Autosampler Manager”, if not connected yet, and
act on the “Virtual Screen” tab to exit from the SETUP mode.
While in SETUP mode, no operations (start of sample sequence, syringe
wash) are allowed.
A non-blocking error has occurred. Plunger motor is selected.
Keypad enters in error modality:
•
by acting on the “Upward arrow” / “Downward arrow” key you
can move the selected motor
•
by pressing the “MOVE” key you can select a different motor
•
by pressing the “STOP” once to retry the operations that have
generated the error.
•
by pressing the “STOP” twice to abort the operations that have
generated the error.
A non-blocking error has occurred. Needle motor is selected.
Keypad enters in error modality:
•
by acting on the “Upward arrow” / “Downward arrow” key you
can move the selected motor
•
by pressing the “MOVE” key you can select a different motor
•
by pressing the “STOP” once to retry the operations that have
generated the error.
•
by pressing the “STOP” twice to abort the operations that have
generated the error.
A non-blocking error has occurred. Turret motor is selected.
Keypad enters in error modality:
•
by acting on the “Upward arrow” / “Downward arrow” key you
can move the selected motor
•
by pressing the “MOVE” key you can select a different motor
•
by pressing the “STOP” once to retry the operations that have
generated the error.
•
by pressing the “STOP” twice to abort the operations that have
generated the error.