HT2000H_en_n.doc Page 126 of 167
Example: only one injector is defined in setup, but “Rear” or “Confirmation” injection mode is used
in the sequence step.
Solution:
Tap “EXIT”. Modify the injection mode in the sequence step (see paragraph 5.3.2 “Setting
sequence step parameters“) or the injector type/number in setup setup (see paragraph 4.2.1
“Setup: Alignment “). In the latter case run the injector alignment.
Restart the automatic run.
8.9 Missing errors
8.9.1 Missing: front injector alignment
Error message:
“Missing: front injector alignment”
Causes:
Front injector alignment (4.2.1.2 “Alignment: Front Injector”) has not been performed after the
change of Analyzer brand or model (4.2.1.1 “Alignment: Analyzer Default”).
Solution:
Perform the Front injector alignment (4.2.1.2 “Alignment: Front Injector”).
8.9.2 Missing: rear injector alignment
Error message:
“Missing: rear injector alignment”
Causes:
Rear injector alignment (4.2.1.3 “Alignment: Rear Injector”) has not been performed after the
change of Analyzer brand or model (4.2.1.1 “Alignment: Analyzer Default”).
Solution:
Perform the Rear injector alignment (4.2.1.3 “Alignment: Rear Injector”).
8.9.3 Missing: touch & plunger zero
Error message:
“Missing: touch & plunger zero”
Causes:
“Touch Plunger Zero” procedure is missing (“4.2.1.4 “Alignment: Touch Plunger Zero”).
Solution:
Execute the “Touch Plunger Zero” procedure (“4.2.1.4 “Alignment: Touch Plunger Zero”).