HT2000H_en_n.doc Page 125 of 167
Solution:
Tap “EXIT” and/or:
edit the syringe volume set in the method;
replace the syringe with one that has the correct capacity;
change the syringe volume set during installation if this doesn't match the installed syringe.
Then restart the automatic run.
8.8.2.2 Tray Type mismatch
Error message:
“Warning: Run Error parameters” “Step X: Tray Type mismatch” (Where “X” is the sequence step)
Causes:
The tray type set in the “TRAY TYPE” screen (see paragraph 7.6.1.1 “Rack mounting”) is different
from that set in the sequence step (see paragraph 5.3.2 “Setting sequence step parameters“).
Solution:
Tap “EXIT” and/or:
• change the tray type set in the sequence step to match the installed tray (see paragraph
5.3.2 “Setting sequence step parameters“);
• change the tray type set in the “TRAY TYPE” screen (see paragraph 7.6.1 “Change
tray/rack type”) to match the installed tray.
• replace installed tray with the correct one (see paragraph 4.2.1.5 “Alignment: Tray Install”)
and change the tray type set in the “TRAY TYPE” screen (see paragraph 7.6.1 “Change
tray/rack type”).
Then restart the automatic run.
8.8.2.3 Injection Mode mismatch
Error message:
“Warning: Run Error parameters” “Step X: Injection Mode mismatch”
Causes:
The injection mode in the sequence step (see paragraph 5.3.2 “Setting sequence step
parameters“) is different from the injector type/number in the setup (see paragraph 4.2.1 “Setup:
Alignment “).