HT2000H_en_n.doc Page 65 of 167
3.5.2 Autosampler connected directly to the PC
3.5.2.1 Autosampler connected directly to the PC by the Ethernet port
1) Connect the autosampler directly to the PC, inserting the Ethernet cable provided with the
autosampler in the port indicated with the number 4 in Figure 23 and then into the Ethernet
port of the PC (see paragraph 3.3.3 “Electrical connections”);
2) The IT/ network administrator must verify that the PC you are working with has a static IP
address (do not work with DHCP);
3) Then operate as described in paragraph 3.5.1.2 “Operations“ starting from point 3).
3.5.2.2 Autosampler connected directly to the PC by a RS232 port
1) Connect the autosampler directly to the PC, inserting the serial cable provided with the
autosampler in the port indicated with the number 4 in Figure 22 and then onto the PC in
any RS232C port available. (see paragraph 3.3.3 “Electrical connections”);
2) Then operate as described in paragraph 3.5.1.2 “Operations“ starting from point 3). The
only difference is that, instead of selecting “TCP/IP port” (see point 9)), you have to select
“Serial Port” and then choose the product model and the correct communication port from
the combo boxes.