8.8 Mismatch errors
8.8.1 Single injection mismatch error
If the method parameters conflict with those defined in “setup”, the unit will display an error
message.
Examples are shown below.
Syringe Volume mismatch
Error message:
“Warning: Run Error Parameters” “Method X: Syringe Volume mismatch”
(Where “X” is the method used for the single injection)
Causes:
The syringe volume set during installation (see paragraph 4.2.2 “Setup: Syringe warmer assembly
installation “) is different from the syringe volume set in the method (5.2.2.1 “General parameters”).
Solution:
Tap “EXIT” and/or:
edit the syringe volume set in the method;
replace the syringe with one that has the correct capacity;
change the syringe volume set during installation if this doesn't match the installed syringe.
Restart the single injection.
8.8.2 Automatic run mismatch errors
If the method parameters in any part of an automatic run conflict with those defined in “setup”, the
unit will display an error message. The unit will show where the problem occurs. These include the
following:
8.8.2.1 Syringe Volume mismatch
Error message:
“Warning: Run Error Parameters”
“Step X: Syringe Volume mismatch in Method Y”
(Where “X” is the sequence step and “Y” is the method used)
Causes:
The syringe volume set during installation (see paragraph 4.2.2 “Setup: Syringe warmer assembly
installation “) is different from the syringe volume set in the method used (5.2.2.1 “General
parameters”).