HT2000H_en_n.doc Page 114 of 167
“MENU” to display method or sequence information; to inject the sample immediately (see
paragraph 7.3.2 “Immediate Injection “), to skip the sample injection (see paragraph 7.3.3 “Skip
Next Vial option “), or to add or remove samples in the tray/s; “STOP” to end the automatic run (see
paragraph 7.3.1 “Interrupting an automatic run).
Select “Exit” to exit from the “Automatic Run” screen and return to the home screen.
7.3.1 Interrupting an automatic run
During the automatic run, from the “AUTOMATIC EXECUTION SCREEN” (see paragraph 7.3
“Automatic run”), tap “STOP” to stop the procedure, then:
• tap “ABORT” to stop the unit immediately;
• tap “TERMINATE OVEN” to continue the execution of the samples present in the oven;
• tap “RESUME” to continue operations as before.
7.3.2 Immediate Injection
During the automatic run, the “AUTOMATIC EXECUTION SCREEN” is displayed (see paragraph
7.3 “Automatic run“). Then follow the access sequence: AUTOMATIC EXECUTION
SCREEN>MENU>IMMEDIATE INJECTION
This function can only be activated during the sample conditioning. When this option is activated
the autosampler stops the current sample conditioning and immediately injects it.
Once the immediate injection has been made, the autosampler will re-start the automatic injection
procedure from the point of interruption.
7.3.3 Skip Next Vial option
During the automatic run, the “AUTOMATIC EXECUTION SCREEN” is displayed (see paragraph
7.3 “Automatic run“). Then follow the access sequence: AUTOMATIC EXECUTION
SCREEN>MENU>SKIP NEXT VIAL
This function can only be activated during the sample conditioning. When this option is activated,
the autosampler stops the current sample conditioning and replaces the sample to its original
position of the tray without injecting it.
The autosampler will re-start the automatic injection procedure from the point of interruption.