HT2000H_en_n.doc Page 15 of 167
Method development option
During method development it can be useful to test methods characterized by a progressive
increment of a specific parameter in order to define which are the best settings for this parameter.
The parameters that can be increased in this way are the syringe/oven temperature and the
incubation time.
This option is only available with “HTA Autosampler Manager”, using the “Method development”
tab. For the detailed explanation of this option, please refer to “HTA Autosampler Manager” online
help.
1.7 Versions
1.7.1 Description
The autosampler is available in the following versions:
HT2000H
This version is equipped with a touch screen interface.
It has a moving tray with 42 vial positions.
It has a 6 position oven (incubation temperature up to 170°C) equipped with an orbital shaker
HT2100H
This version is equipped with a keypad.
It has a fixed tray with 14 vial positions.
It has 1 position oven (incubation temperature up to 150°C) with a sussultatory shaker.
Require a PC for set up, service, method and sequence editing.
HT2000HT
This version is equipped with a touch screen interface.
It has a moving tray with 42 vial positions.
It has a 3 position oven (incubation temperature up to 300°C) equipped with an orbital shaker.
1.7.2 Options
Available for all versions:
1.90.501 - HTA Autosampler Manager (Standard Version)
PC software. This may be sold as an option. Please check with your service representative.
1.90.502 - HTA Autosampler Manager (CFR 21 Part 11)
PC software compliant with CFR 21 Part 11 regulation. This may be sold as an option. Please
check with your service representative.
1.91.295 - System integrity kit
The System integrity kit is sold as an option. The System integrity test is used to verify the syringe
integrity. This test can easily be executed at the beginning of each batch in a very convenient way.