HT2000H_en_n.doc Page 88 of 167
l) Only for HT2000HT. If Syringe Warmer High Temperature Assembly for HT2000HT has
been installed you should select as syringe type: “High Temperature”. If Syringe Warmer
Assembly for standard version syringe has been installed you should select as syringe
type: “Standard”. If the screen for the syringe type selection does not appear contact the
Technical Assistance Service.
m) The autosampler will start “Plunger zero” to automatically complete syringe installation.
n) The autosampler will also automatically start the syringe conditioning. This procedure
should take about 90 minutes.
Warning
Set-up must be completed by competent personnel, properly trained regarding the procedures
described in this manual: any missing or different execution of the described procedures can
cause damage to the autosampler or to the person operating it.
However, in particular, you should pay attention to the moving parts such as the sledge, syringe,
needle, the cover of the oven etc...Some parts of the autosampler (oven, syringe, needle or
vials) can work at high temperatures. Be careful to avoid burns. Allow these parts to cool down
before handling them.
Wrong settings could cause damage to the syringe and/or to the autosampler.
Warning
The syringe conditioning procedure can be skipped during the autosampler installation because
the syringe has been already conditioned by the manufacturer. This procedure can also be
skipped in case of the same syringe re-installation. If the syringe barrel or the syringe plunger or
both the parts are new, the procedure must be carried out.
4.2.2.1 Syringe warmer assembly removal
Warning