HT2000H_en_n.doc Page 93 of 167
• Maintenance: displays information and changes settings about preventive maintenance
counters (see paragraph 4.2.4.5 “Setup configuration: Maintenance“);
• Advanced: edits the advanced parameters of the autosampler (see paragraph 4.2.4.6 “Setup
configuration: Advanced”);
• Activation: enables the optional features of the autosampler (see paragraph 4.2.4.7 “Setup
configuration: Activation”).
4.2.4.1 Setup configuration: Sound
Access sequence: RUN>SETTINGS>SETUP>CONFIGURATION>SOUND
• Select “2” to emit a sound when the touch screen is tapped;
• Select “1” to emit a sound after the end of each single injection, every sample during an
automatic run, when “save” or “store” is tapped, or if the autosampler memory is erased;
• Select “0” to disable the sound emission.
Tap “SAVE” to store the modification, otherwise “CANCEL”.
Select “Exit” to exit from this screen and return to the “Setup Configuration”” screen (see paragraph
4.2.4 “Setup: Configuration“).
4.2.4.2 Setup configuration: Light
Access sequence: RUN>SETTINGS>SETUP>CONFIGURATION>LIGHT
This is used to change the light setting
• Select “On” to switch on the syringe illumination;
• Select “Off” to switch off the syringe illumination;
• Select: “Injection” to switch on the syringe illumination during injection only; Tap “SAVE” to store
the modification, or “CANCEL”.
Select “Exit” to exit from this screen and return to the “Setup Configuration” screen (see paragraph
4.2.4 “Setup: Configuration”).
4.2.4.3 Setup configuration: Run
Access sequence: RUN>SETTINGS>SETUP>CONFIGURATION>RUN
Use this to change the following operational parameter:
• Tray Behaviour: to set tray position during automatic run;
Tray
Behaviour
Closed: the tray remains closed during run
Open: the tray remains open during run
This parameter is not editable for the HT2100Hversion.
• Power-on Restart: sets autosampler restart after a power failure. It should be set to “ON” only if
your GC/analyzer and your data acquisition system are also able to restart automatically;