HT2000H_en_n.doc Page 79 of 167
-If the vial locator does not find the injector, the following message appears: “Injector not found”. In
this case, only the “DO MANUAL” or “ABORT” buttons are enabled.
Select “Exit” to exit from this screen and return to the “ALIGNMENT” screen (see paragraph
4.2.1 “Setup: Alignment“).
4.2.1.3 Alignment: Rear Injector
Access sequence: RUN>SETTINGS>SETUP>ALIGNMENT>REAR INJECTOR
Tap “Rear Injector” from the “Setup: Alignment” screen to set the alignment of the rear injection
(see paragraph 4.2.1 “Setup: Alignment”).
The procedure for rear injector alignment is the same as for the front injector (see paragraph
above).
Select Exit to exit from this screen and return to the ALIGNMENT” screen (see paragraph
4.2.1 “Setup: Alignment”).
4.2.1.4 Alignment: Touch Plunger Zero
Access sequence: RUN>SETTINGS>SETUP>ALIGNMENT>TOUCH PLUNGER ZERO
Follow the instructions of the wizard. Place a capped sample vial in the indicated position and
ensure the capped waste vial is in the waste position. If the System integrity kit has been provided,
also place the System integrity tool in the position indicated with 6 in Figure 1 for
HT2000H/HT2000HT and Figure 9 for HT2100H. Tap “Continue” to start, (or “ABORT” to quit). The
plunger touch procedure is then run automatically. The vial locator touches the indicated position of
the oven cover, the indicated vial in the oven, the indicated vial in the tray, the injector/s and on the
System integrity area. Then the syringe zero procedure is performed. This procedure can be
stopped or skipped by tapping the appropriate button.
Select Exit to exit from this screen and return to the ALIGNMENT” screen (see paragraph
4.2.1 “Setup: Alignment”);
4.2.1.5 Alignment: Tray Install
Access sequence: RUN>SETTINGS>SETUP>ALIGNMENT>TRAY INSTALL
This screen shows a list of trays/racks stored in memory. The following options are available:
Add new: adds a new tray/rack;
Remove: removes a tray/rack;
Select a tray to view or edit its details. The number of vial positions, dimension of the sample vials
and the needle draw depth inside the sample vial can be modified if your sampler allows it. If you
modify a parameter, a screen displays: “Modifying tray”. Tap “SAVE to confirm, otherwise tap
“CANCEL”. A confirmation screen appears “Tray parameters has been modified”. Tap
“CONTINUE“.