10–27
Transpector CPM Operating Manual
10.6.2.5.2 Time Between Scans
Time Between Scans can be set to a value greater than the minimum scan time
if additional wait time is desired. This field can only be set if Manual is chosen as
the Delay Mode.
Figure 10-47 Time between scans
NOTE: Time Between Scans cannot be set lower than the minimum scan time.
10.6.2.5.3 Stabilization Time
Stabilization Time is the amount of time FabGuard Explorer waits after the Start
Mode condition has been met before setting the Emission and EM to their specified
Start states and beginning data acquisition. For example, if the recipe starts based
on an analog input and an additional wait time is desired before the Emission is
turned on.
10.6.2.6 Correction
Defines the type of Baseline correction used by FabGuard Explorer (if any
correction at all) and whether or not Peak Lock is turned on.
10.6.2.6.1 Baseline
Sets the Baseline to Off or Mono. For more information, refer to the FabGuard
Explorer Operating Manual.
Figure 10-48 Baseline