EasyManua.ls Logo

Inficon Transpector CPM - Spectrum Mode Acquisition Parameters

Inficon Transpector CPM
190 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
10 14
Transpector CPM Operating Manual
10.6.1.1.1 Spectrum Mode Acquisition Parameters
Figure 10-22 Spectrum mode acquisition parameters
Start Mass (amu) . . . . . . . . . . . . . . The amu at which the RGA will
begin scanning.
Stop Mass (amu). . . . . . . . . . . . . . . The amu at which the RGA will
stop scanning.
Points per amu . . . . . . . . . . . . . . . . The number of points per amu (ppamu) that
the RGA will monitor.
Setting a Stop Mass value outside of the sensor’s 2222range will create an error.
For example, if a 200 amu sensor is installed, FabGuard Explorer will not allow a
250 amu Stop Mass.
Increasing the number of ppamu will increase the total time it takes to perform a full
scan. In FabGuard Explorer, there are four different options for ppamu values: 1, 2,
5 and 10 ppamu.

Table of Contents