7–18
Transpector CPM Operating Manual
2 Measure the resistance of each of the pins with respect to ground
(pin 7 and 8). These measurements must be above 30 M.
3 Measure the resistance of each of the pins with respect to each other. All of
these measurements must be above 30 M, with the exception of the
measurement between the filament pins if the filament has not failed.
7.7.2 Transpector Sensor Filament Replacement
1 Remove the ceramic sealing disk (A) by carefully pulling the wire hold downs
(B) away from the edges of the disk and pulling off the sealing disk.
2 Remove spring (C).
3 Remove the filament lead screw (D1), loosen the filament lead screw (D2) and
remove the filament leads (E) from the filament. (See Figure 7-17.)
Figure 7-17 Sensor head
A
BC D1 ED2