8–2
Transpector CPM Operating Manual
8.3  CPM Symptom—Cause—Remedy Chart
Before proceeding, make sure that the insulation on all cables is intact. There must 
be no damage to the insulating material. If there is, replace the cable
. 
Table 8-1  Symptom—Cause—Remedy Chart
SYMPTOM CAUSE REMEDY
CPM will not pump down Forepump not 
plugged in
Connect cable to controller
Control cable missing or 
bad connection
Check control cable
Foreline isolation valve not 
opening
Check valve actuation
Digital Pressure Switch 
defective or not within 
required pressure range
Replace the Digital 
Pressure Switch
Verify incoming pneumatic 
pressure is between 58 and 
100 psi
Turbo Molecular Pump 
does not reach operating 
speed
Foreline pressure too high 
due to gas leak or 
forepump diaphragm 
failure
Find Leak(s).
Replace diaphragm. (refer 
to section 7.6 on page 7-6)
Turbo Molecular Pump 
bearing drag excessive
Replace pump
Foreline pressure too high 
due to water vapor in 
foreline hose
Run pumpdown and purge 
with dry gas
No pneumatic pressure 
applied to the foreline valve
Check supply and 
connections to the foreline 
block valves
Turbo Molecular Pump is 
noisy
Bearing failure Replace pump
Diaphragm Pump is noisy Loose Diaphragm Tighten Diaphragm
Analyzer pressure high 
with sampling valves 
closed
Leaks in manifold or 
orifices
Tighten or replace gaskets
Seat leaks to process gas Replace seat or valve