If there is a need to increase treatment energy, do so in gradual steps and monitor any
thermal rise after each sonication.
[W-27]
11. The system will detect and notify about movements on the next slice, according to the Treatment
Direction defined in the Treatment Protocol.
12. If required, enter Deformation Mode to update the treatment plan on the Axial Current Anatomy
strip (refer to Section 7.2.10).
13. Repeat the procedure until the entire ROT is covered with Thermal Dose.
14. If required, click Replay to view the results of previous sonications.
7.4.2. Adding a new Sonication Spot
Click this button and add a new sonication spot in the Selected Image window. Spots can be added
on a Current Anatomy image only.
◼ In Align Mode: It is possible to add a Regular spot only. Make sure to add the spot at the center of
the ROT.
◼ In Treat Mode: There are three optional spot types that could be added:
● Macro-spot: Click within the ROT polygon and the system will initiate planning a sub-spots
arrangement that will cover the ROT while taking into consideration all LEDR limitations. Click in
the ROT at the side that you with the Macro-spot to start at.
● Regular Spot: Adding a Regular spot is possible not only within the ROT but within the entire
prostate capsule.
● Macro Sub-spot: Select an existing Macro-spot, then select this option to add a Macro sub-spot
to the selected Macro. The added sub-spot will be sonicated according to its relative rotation
degree.