GETTING STARTED
Manual Part Number
57
CAUTION:
If any of the above inspections or tests fail to meet the expected values, discontinue use
of the system until it has been thoroughly inspected by authorized INSIGHTEC service
personnel.
Do not treat patients if the DQA was not completed successfully!
[C-5]
NOTE:
Exablate Prostate treatments create many new series on the MR scanner. It is advisable
that during DQA, sufficient disk space be available on the MR workstation.
[N-11]
3.8. Storage and Maintenance of the DQA Phantom Gel
The DQA phantom gel is a solid water based cross-linked gel, embedded in a rounded plastic container, and
protected by a Mylar membrane.
1. Always check the expiration date on the DQA phantom gel before using it; If it is expired, order a new
phantom gel. Expired phantom gel should not be used.
2. Inspect the phantom for cracks and chips; if any appear, order a new phantom gel.
3. If the surface of the phantom is defective, a replacement phantom must be ordered from INSIGHTEC
immediately. The defect may trap air and cause beam deflection.
4. Verify that the Mylar membrane protecting the phantom gel is intact and tightened. In any case of holes
in the Mylar membrane, bacteria may quickly destroy the phantom.
5. Close the bottom cover of the phantom gel and store it in a cool, dark location.
NOTE:
The DQA phantom gel is fragile, Handle with care.
[N-12]