HIGH-LEVEL DISINFECTION
Manual Part Number
a. Fill container #1 (“cleaning solution”) with 20 ml of Enzol® enzymatic detergent and then
add approximately 2.5 liters (85 oz) of purified water. Gently shake the container to mix the
solution.
b. Fill container #3 (“disinfectant”) with approximately 2.5 liters (85 oz) of Revital-Ox™.
c. Fill containers #2 and #4 with purified water, approximately 4 liters (135 oz) in each.
C.3.2 Post-Treatment Procedure
1. Remove disposable shell from the probe - pull its balloon “dress” over the inflated part to expose
the locking pin. Release the pin by firmly pulling its nylon cord, and then pull the disposable shell
assembly from the probe. Remove white O-ring from the probe as well.
2. Dispose all single-use items according to the policies of the medical facility.
3. Place the cleaning and disinfection cart adjacent to the Exablate table and attach (screw) the water
fitting connector (previously connected to the water pouch) to the connector in the back of the
cleaning and disinfection cart.
C.3.3 Probe Cleaning
The transducer surface is very delicate. Clean and handle it very gently during the
procedure (e.g., when wiping the probe) and avoid any contact with sharp objects.
[W-41]
1. Press “Clean” on the Exablate table water system display (“display”) and choose “HLD”.
2. Press “Next” on the display (if the cleaning cart preparation, described above, was completed).
3. Verify that the table water fitting connector is properly attached to the connector at the back of
the cleaning and disinfection cart and press “Next” on the display.
4. Use two (2) disinfection wipes (consecutively) to carefully wipe the probe for two (2) minutes, then
press “Next”.
a. If necessary, press “Back” on the display to go to previous screen.
5. Lock probe’s shell and soak the probe in the cleaning solution by inserting it into container #1
(“cleaning solution”) for six (6) minutes, then press “Next”.
6. Remove the probe from container #1, wait until it stops dripping, and soak it in water by inserting
it into container #2 (“water”) for three (3) minutes, and then press “Next”.