A.3.3. Parameter Settings
◼ Pre-Sonication Phase
Use Pre-Sonication Phase to determine the number of “cold phases” created by the system before
each sonication.
◼ Post-Sonication Phase
Use the Post-Sonication Phase to determine the number of cooling phases the system scans after
each sonication.
◼ CV rhrcctrl
This defines the image types reconstructed for each phase. For reconstructing magnitude, real and
imaginary images, use a value of 29.
◼ CV rhtype
For EPI sequences, use a value of 8. This applies only to EPI sequences.
◼ External Trigger CV Name
This defines the name of the external trigger.
◼ External Trigger CV Value
For sequences with the ext_trig CV, use a value of 1.
For sequences without the ext_trig CV, use a value of 0.
◼ Number of Slices
For multi-slices thermal imaging, defines the number of slices per sonication.