ADVANCED OPTION MODE
Manual Part Number
155
A.3.3. Parameter Settings
Pre-Sonication Phase
Use Pre-Sonication Phase to determine the number of “cold phases” created by the system before
each sonication.
Post-Sonication Phase
Use the Post-Sonication Phase to determine the number of cooling phases the system scans after
each sonication.
CV rhrcctrl
This defines the image types reconstructed for each phase. For reconstructing magnitude, real and
imaginary images, use a value of 29.
CV rhtype
For EPI sequences, use a value of 8. This applies only to EPI sequences.
External Trigger CV Name
This defines the name of the external trigger.
External Trigger CV Value
For sequences with the ext_trig CV, use a value of 1.
For sequences without the ext_trig CV, use a value of 0.
Number of Slices
For multi-slices thermal imaging, defines the number of slices per sonication.