HIGH-LEVEL DISINFECTION
Manual Part Number
162
7. Remove probe from container #2 and dry it entirely with a lint free cloth for two (2) minutes, and
then press “Next”.
8. Visually inspect the probe for visible soil/residues and press “Next”.
a. If soil/residues are found - repeat probe cleaning by pressing “Repeat” on the display.
b. If other visible signs of deterioration such as corrosion, discoloration, pitting, or cracks are
noticed, contact Technical Support.
C.3.4 Probe Disinfection
1. Use two (2) disinfection wipes (consecutively) to carefully wipe the probe for two (2) minutes, and
then press “Next”.
2. Dry the entire probe with a lint-free cloth for two (2) minutes (if necessary, continue until visually
dry), and then press “Next”.
3. Verify that the white tray of the cart is placed over container #4.
4. Connect the water hose (on the cleaning and disinfection cart) to container #3 (“disinfectant”), as
shown on display, and then press “Next”.
5. Soak the probe in Revital-Ox™ by inserting it into container #3 (“disinfectant”).
a. Verify that the probe and the water hose are properly connected to container #3.
6. Press “Start” to begin circulation and wait for system to finish.
a. The system will circulate automatically for one (1) minute.
b. If necessary, press “Stop” to stop circulation (to resume press “Start” again).
7. When requested, confirm that the liquid level in container #3 is up to its upper marker. Add
disinfectant if needed.
a. To quit HLD, press “Home” and confirm pop-up message to return to “Home” screen.
8. Press “Start” to begin circulation and wait for system to finish.
a. The system will circulate automatically for approximately thirteen (13) minutes.
b. If necessary, press “Pause” to pause circulation.
c. When circulation is paused, it is possible to do one of the following:
i. Resume circulation by pressing “Start”.
ii. Return to previous screen by pressing “Back”.
iii. Go back to “Home” screen by pressing “Home” and confirming the pop-up
message.