PLANNING STAGE
Manual Part Number
98
WARNING:
The dose transfer procedure refers to the prostate capsule and the ROT as rigid structures,
and thus may fail to achieve accurate results in case of significant deformations. In such
cases, delete the previously prostate capsule and ROT objects, and draw new ones.
[W-19]
5. Rectal wall, balloon boundaries and all LEDR drawings were deleted. Re-draw the rectal wall and all
necessary LEDR contours, and detect balloon boundaries before continuing (see Section 6.7).
6. Proceed to Treatment stage to resume the treatment. The system will ask whether to keep the
accumulated dose or to delete it.
WARNING:
Carefully review the dose transfer procedure result before deciding whether to keep or to
delete the accumulated dose. Inaccurate adjustment of the accumulated dose overlay
may be misleading when attempting to estimate the region of treatment dose coverage.
[W-20]
Figure 6-10: Dose Transfer procedure