Improper use of the Advanced Option Mode may degrade the quality of treatment and
may even result in personal injury.
[W-39]
Advanced Options Mode enables the operator to override the automatically generated treatment plan and
manually set treatment parameters. This includes:
◼ Sonication parameters
◼ MR scan protocol parameters
When you access the Advanced Options dialog box, all displayed values are the current treatment values.
To change a parameter:
1. Click the button next to the parameter to be changed or adjusted; the parameter’s value text box is
activated.
2. Type in or select the required value.
3. Click on:
◼ OK to accept the changes
◼ Cancel to reject all changes and close the dialog box
◼ Reset Values to reset all changes
To return the parameter to system default:
1. Open the Advanced Options dialog box.
2. Click the box next to the parameter; the value text box is disabled, and the value is automatically
generated by the system.