HIGH-LEVEL DISINFECTION
Manual Part Number
163
NOTE
If the temperature of the water in the system is below threshold, the system will first go
through a warming stage (“Warming” will appear on top of the display), and only when
completed, will the system automatically move to circulation.
[N-53]
C.3.5 Probe Rinsing
1. After probe disinfection is successfully completed, carefully disconnect both the water hose and
the probe from container #3 and wait until the probe stops dripping.
2. Move the white tray of the cart over container #3 and then connect both the water hose and the
probe to container #4 (“water”).
a. To quit HLD, press “Home” and confirm pop-up message to return to “Home” screen.
b. Verify that the probe and the water hose are properly connected to container #4.
3. Press “Start” to begin circulation and wait for system to finish.
a. The system will circulate automatically for two and a half (2.5) minutes.
b. If necessary, press “Pause” to pause circulation.
c. When circulation is paused, it is possible to do one of the following:
i. Resume circulation by pressing “Start”.
ii. Go back to “Home” screen by pressing “Home” and confirming the pop-up message.
C.3.6 Post-HLD
1. After probe rinsing is completed, disconnect both the water hose and the probe from container #4,
then press “Next”.
2. Dry the entire probe with three (3) lint-free cloth for three (3) minutes, and then press “Next”.
3. Cover the probe with probe storage cover and secure with the cover label.
4. Then, put the mechanical protective cover (P/N MEC501226) over the transducer (probe tip) and
press “Next”.
5. Disconnect the Exablate table water fitting connector from the connector in the back of the
cleaning and disinfection cart.
6. Dispose of the cleaning and disinfection materials (contents of the 4 containers) according to the
respective manufacturer instructions and the medical facility policy.