HIGH-LEVEL DISINFECTION
Manual Part Number
164
7. Completely fill the four (4) containers with Sodium hypochlorite (5.25-6.15%) diluted approximately
1:100 with potable water (e.g., 25 cc of Sodium hypochlorite for 2.5 liters of water or 40 cc of
Sodium hypochlorite for 4 liters of water).
8. Wait for five (5) minutes and then completely drain the containers.
9. Rinse the containers, twice, by filling them with potable water and then drain completely.
10. Dry all the containers and caps with a lint-free cloth.
11. Allow the containers and caps to air-dry before storing them.
12. Press “Finish” on the display and confirm pop-up message to finish and exit HLD procedure.
13. If HLD procedure is completed successfully, an approval message will appear on display.
a. Press “OK” to return to “Home” screen.
14. If the HLD procedure was not completed successfully, a warning message will appear on the display.
a. Press “OK” to return to “Home” screen.
15. Continue to system shutdown, as described in Operator’s Manual, or to probe and table preparation
for treatment (only if HLD completed successfully).
WARNING
If HLD procedure failed to be completed successfully do not treat patients and contact
technical support at TechSupp@insightec.com, or by calling your local
service/applications team!
[W-42]
WARNING
After the issue is resolved by Technical Support, perform full HLD procedure successfully,
before treating any patient.
[W-43]
WARNING
After six (6) months, contact Technical Support for maintenance.
[W-44]